Stryker Hospital Bed Motor Repair

Manufacturer:
Stryker
Application:
Actuator Repair,Medical Equipment,
Description:

Stryker Hospital Bed Motor Repair

Hospital Bed Motor Rebuild includes the following:


Motor disassembly

Windings testing:  test for shorted windings and electrical resistance ohm test.  Test for insulation breakdown by 1,000 volt hi-pot, and a 500 volt megger test.

If the stator windings do not pass the above tests, the stator is rewound to original specifications.

Rotor is check for proper balance, shaft straightness, and bearing surface wear. 

Motor bearing journals are inspected and machined if needed

Gearbox is inspected, cleaned, and re-greased.  Shafts are inspected for straightness and wear, Gears are inspected for proper pitch and excessive wear.

Bearings

Capacitor

Motor reassembly and testing

Cleaning and painting

Labor

Warranty

 

If a motor requires more than the normal rebuild, or is beyond economical repair, we will notify you immediately and await your instructions before proceeding.

 

Please package your motor carefully and send it to us along with a copy of this webpage.